Description

  • Simulation by: Martin Prosi
  • Added: Saturday, June 19 2004
  • References: N/A
  • Comment: Restriction of the pulmonary artery for complex congenital heart defect. Comparison of the flow fields between the Flowatch device and circular banding. Quasi-stationary flow field (Re=500).The computational results demonstrate that the complex flow patterns occur downstream the non-axisymmetric FloWatch constriction. The results also demonstrate that the pressure drop accross the banding mainly depends on the cross sectional area and not on the shape of the banding.
  • Time stepping: 400 time steps of 0.01s
  • Navier-Stokes solver: velocity-pressure splitting method with algrbraic pressure-corrected preconditioner.